TB 6-6625-2329-24
(4) Press TI PARAMETER LEVEL pushbutton, enter DATA numerical value,
and press ENTER/UNITS dBm key for each row in table 5. Measure RF power using
measuring receiver. If measuring receiver indications are not within limits specified in
Test instrument
Measuring receiver
DATA and
indications
ENTER/UNITS
(dBm)
(dBm)
Min
Max
13
11.5
14.5
0
-1.5
1.5
(5) Press TI PARAMETERRF pushbutton, enter DATA numerical value, and
press ENTER/UNITS frequency key for each row in table 6. Measure and record the RF
power using the measuring receiver.
(6) Calculate the flatness using the formula below. If calculated flatness is not less
than the maximum limits listed in table 6, perform b below.
FORMULA
Flatness = (highest reading lowest reading)/2
Measuring receiver
Test instrument
indications
DATA and
ENTER/UNITS
MHz
Reading
Maximum limit
(dB)
<(dB)
.1
1
.2
1
.3
1
.4
1
.5
1
.6
1
.7
1
.8
1
.9
1
1
1
2
1
3
1
4
1
5
1
6
1
9