TB 9-6625-2329-35
(2) Set up frequency counter controls to measure frequency with 50 Ω input.
(3) Press pushbuttons as listed in (a) through (f) below:
SHIFT-INSTR PRESET.
(a)
PARAMETER-RF.
(b)
ON/OFF-EXT AC.
(c)
PARAMETER-LEVEL.
(d)
DATA-0.
(e)
ENTER/UNITS-dBm.
(f)
(4) Press TI PARAMETER-RF pushbutton, enter DATA numerical value, and
press ENTER/UNITS MHz key for each row in table 3. Frequency counter will indicate
within the limits specified.
Test instrument
Frequency counter indications
DATA and
(MHz)
ENTER/UNITS
(MHz)
Min
Max
10
9.999999
10.000001
60
59.999999
60.000001
90
89.999999
90.000001
150
149.999999
150.000001
(5) Press PARAMETER-LEVEL pushbutton and press ON/OFF-OFF key.
(6) Disconnect TI RF 50Ω output at frequency counter A input.
(7) Connect TI RF 50Ω output to frequency counter C input.
(8) Set frequency counter to C input.
(9) Press PARAMETER-LEVEL pushbutton and press ON/OFF-INT ON key.
8