TB 9-6625-2329-35
(10) Press TI PARAMETER-RF, enter DATA numerical value, and
ENTER/UNITS MHz key
for each row in
table 4. Frequency will
indicate within
limits
specified.
Test instrument
DATA and
(MHz)
ENTER/UNITS
Min
Max
(MHz)
400
399.999999
400.000001
700
699.999999
700.000001
1000
999.999999
1000.000001
(11) Press PARAMETER-LEVEL pushbutton and press ON/OFF-OFF key.
(12) Disconnect TI RF 50Ω from frequency counter.
b. Adjustments. No adjustments can be made.
9. RF Output and
Flatness
a. Performance Check
(1) Set measuring receiver pushbuttons as listed in (a) through (e) below:
(a) Connect sensor module to measuring receiver. Zero, calibrate, and save
sensor values as necessary.
(b) INSTR PRESET.
(c) LOG/LIN to LOG.
(d) Enter 32.0 SPCL (0.01 dB resolution).
(e) Connect sensor module input to TI RF 50Ω.
(2) Press pushbuttons as listed in (a) through (d) below:
(a) SHIFT-INSTR PRESET.
(b) PARAMETER-RF.
(c) DATA-100.
(d) ENTER/UNITS-MHz.
(3) Press TI PARAMETER-LEVEL pushbutton, enter DATA numerical value,
and press ENTER/UNITS dBm key for each row in table 5. Measure RF power using
measuring receiver. If measuring receiver indications are not within limits specified in
Test instrument
Measuring receiver
DATA and
indications
ENTER/UNITS
(dBm)
(dBm)
Min
Max
13
11.5
14.5
0
-1.5
1.5