TB 9-6625-2329-35
(b) PARAMETER-LEVEL, enter DATA 0 numerical value, and press
ENTER/UNITS dBm key.
(c) PARAMETER-AF, enter DATA 1 numerical value, and press
ENTER/UNITS kHz key.
(d) PARAMETER-AM, enter DATA 80 numerical value, and press
ENTER/UNITS % key.
(e) PARAMETER-AM and press ON/OFF-INT/ON key.
(3) Press pushbuttons on measuring receiver as listed in (a) through (f) below:
(a)
SHIFTINSTR PRESET.
(b)
AM
(c)
PEAK +.
(d)
300 Hz high pass filter.
(e)
15 kHz low pass filter.
(f)
(4) Adjust R503 (fig 4) for a 80 1% indication (R).
(5) Press PARAMETER-LEVEL pushbutton and press ON/OFF-OFF key.
a. Performance Check
(2) Press SHIFT-INSTR PRESET pushbutton.
(3) Press pushbuttons on measuring receiver as listed in (a) through (e) below:
SHIFT-INSTR PRESET.
(a)
ΦM.
(b)
(c)
PEAK +.
(d)
300 Hz high pass filter.