TB 9-6625-2332-50
(8) Using multimeter, measure ohms. Multimeter indication will be within the
limits specified in table 3. Record the measured value as actual termination resistance in
ohms.
Multimeter indication
(Ω)
Min
Max
45
55
(9) Calculate the termination correction factor using the formula below:
[ .5 ] / [ actual termination resistance / ( actual termination resistance + 50 ) )]
example
[ .5 ] / [ 49 / ( 49 +50) )]
[ .5 ] / [ 49 / ( 99 )]
[ .5 ] / [ .4949] = 1.01 = Termination correction factor
(10) Record the termination correction factor in table 4, 5, 6, and 7.