Table 17. Frequency Span Accuracy
Table 18. Frequency Response

TB-9-6625-2333-35 Spectrum Analyzer Agilent Models 8562EC-104 and 8562EC-104 Manual
Page Navigation
  25    26    27    28    29  30  31    32    33    34    35  
TB 9-6625-2333-35
Figure 5. Frequency response - equipment setup.
(2) Press TI keys and enter values using DATA keys as listed in (a) through
(e) below:
(a) INSTRUMENT STATE PRESET.
(b) FREQUENCY then [CENTER FREQ] to 300 MHz.
(c) SPAN then [SPAN] to 1 MHz.
(d) AMPLITUDE then [LOG dB/DIV] to 2 dB/DIV.
(e) [MORE 1 OF 3] then [COUPLING AC DC] to select DC.
(3) Set signal generator frequency to 300 MHz and level output to -5 dBm.
(4) Press TI MARKER PEAK SEARCH key.
(5) Set measuring receiver to measure RF power in dBm at 300 MHz.
(6) Adjust signal generator level output for a TI MKR amplitude indication
between -9.95 and -10.05 dBm.
(7) Record measuring receiver indication.
(8) Set signal generator frequency to 13.2 GHz.
(9) Set measuring receiver to measure RF power in dBm at signal generator
frequency.
(10) Press TI FREQUENCY key then [CENTER FREQ] key and enter 13.2 GHz
using DATA keys.
(11) Press MARKER PEAK SEARCH key.
(12) Adjust signal generator level output for a TI MKR amplitude indication
between -9.95 and -10.05 dBm.
(13) Measuring receiver indication will be within 2.9 dB of indication recorded in
(7) above.