TB 9-6625-2335-35
Test
instrument
parameters
Performance specifications
(% of reading + No. of counts)1, 2
Ac volts true
Accuracy:
rms
Range
30 mV to 30 V
300 V
20
to
45 Hz
0.6 + 1320
0.66 + 1320
45
to
100 Hz
0.25 + 1320
0.31 + 1320
100 Hz
to
20 kHz
0.17 + 1320
0.23 + 1320
400 Hz
to
20 kHz3
0.18 + 1320
0.24 + 1320
20
to
100 kHz
0.7 + 2300
1.1 + 3900
100
to
300 kHz
3.2 + 9900
---
300 kHz
to
1 MHz
10.2 + 66600
---
(% of reading + No. of counts)
Ac current
Accuracy:
1, 2
true rms
Range
30 mA and 300 mA
1A
20
to
45 Hz
0.93 + 2800
1.03 + 2800
46
to
100 Hz
0.38 + 2800
0.48 + 2800
101 Hz
to
20 kHz
0.33 + 2800
0.43 + 2800
400 Hz
to
20 kHz3
0.33 + 2800
0.43 + 2800
21
to
100 kHz
1.08 + 4000
---
Accuracy: (% of reading + No. of counts)1
Range
4-wire Ω
30Ω
0.0075 + 335
300Ω
0.0055 + 35
3 kΩ
0.005 + 7
30 kΩ
0.005 + 7
300 kΩ
0.005 + 8
3 MΩ
0.0065 + 14
30 M Ω
0.04 + 83
Accuracy: (% of reading + No. of counts)1
Range
2-wire Ω
30Ω
0.0075 + 20335
300Ω
0.0055 + 2035
3 kΩ
0.005 + 207
30 kΩ
0.005 + 27
300 kΩ
0.005 + 10
3 MΩ
0.0065 + 14
30 M Ω
0.04
+ 83
Accuracy: (% of reading)
Range:
10
to
400
Hz
0.05
400 Hz
to
1.5 MHz
0.01
1Auto
zero on, 10 PLC, 6-1/2 digits, 1-year specifications.
2Ac coupled, ac slow filter on (ACBAND20).
3Ac fast filter (ACBAND400).
4Checked at 1 kHz only.. Other frequencies checked in ACV function.
3