TB 9-6625-2335-35
Test
Calibrator output settings
Test instrument
instrument
indications
ranges
Volts
Min
Max
30
mV
30 mV
20 Hz
29.8068
mV
30.1932
mV
30
mV
30 mV
50 Hz
29.9118
mV
30.0882
mV
30
mV
30 mV
1 kHz
29.9358
mV
30.0642
mV
300
mV
300 mV
1 kHz
299.358
mV
300.642
mV
300
mV
300 mV
50 Hz
299.118
mV
300.882
mV
300
mV
300 mV
20 Hz
298.068
mV
301.932
mV
3
V
3V
20 Hz
2.98068
V
3.01932
V
3
V
3V
50 Hz
2.99118
V
3.00882
V
3
V
3V
1 kHz
2.99358
V
3.00642
V
30
V
30 V
1 kHz
29.9358
V
30.0642
V
30
V
30 V
50 Hz
29.9118
V
30.0882
V
30
V
30 V
20 Hz
29.8068
V
30.1932
V
300
V
300 V
40 Hz
297.888
V
302.112
V
300
V
300 V
50 Hz
298.953
V
301.047
V
300
V
300 V
1 kHz
299.178
V
300.822
V
(4) Press keys as listed in (a) through (d) below:
(a) Blue shift then CONFIGURATION A/NPLC.
(b) SCROLL ↓ until ACBAND is displayed.
(c) MATH 500.
(d) ENT.
(5) Set calibrator output for settings and press TI RANGE ↑ or ↓ key to select
ranges listed in table 6. TI will indicate within limits specified.
Calibrator output settings
Test
Test instrument
instrument
indications
ranges
Volts
Min
Max
300 V
90 kHz
300 V
296.31
V
303.69
V
30 V
10 kHz
30 V
29.9328 V
30.0672
V
30 V
90 kHz
30 V
29.767
V
30.233
V
30 V
250 kHz
30 V
28.941
V
31.059
V
30 V
500 kHz
30 V
26.274
V
33.726
V
3V
1 MHz
3V
2.6274 V
3.3726
V
3V
250 kHz
3V
2.8941 V
3.0233
V
3V
90 kHz
3V
2.9767 V
3.0233
V
3V
10 kHz
3V
2.99328 V
3.00672
V
300 mV
10 kHz
300 mV
299.328
mV
300.672
mV
300 mV
90 kHz
300 mV
297.67
mV
302.33
mV
300 mV
250 kHz
300 mV
289.41
mV
310.59
mV
300 mV
1 MHz
300 mV
262.74
mV
337.26
mV
30 mV
1 MHz
30 mV
26.274
mV
33.726
mV