TB 9-6625-2337-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
VAR LEVEL:
Amplitude:
0.0 to 5 V
Pulse Width:
.025 to 2.0 μs
Rise Time:
0.1 μs
Fall Time:
0.15 μs
G
75
TC Long
modulation:
Width:
0.5 0.1 μs
Spacing:
15, 30, 60, 120, 240, and 480 μs 5% from Mode 4
GTC trigger or from P3 pulse of an SIF challenge.
GTC Short
modulation:
Width:
0.5 0.1 μs
Spacing:
15, 65, 115 and 165 μs 5% from Mode 4 GTC
trigger or from P3 pulse of an SIF challenge.
MEASUREMENT:
Pulse Width
and pulse
Accuracy:
1 count or 0.025 μs whichever is greater
spacing:
Counter:
Range:
Main & Aux RF I/O: 960 to 1200 MHz
Low Power in:
12 to 1200 MHz
Accuracy: 10 kHz or 0.02%, whichever is greater
Power:
Range:
Main & Aux RF I/O: 25 W or less to 75 W CW
25 W or less to 10 kW Pulsed
Low Power in: 10 mW to 25W CW or Pulsed
Accuracy:
0.5 dB
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are required. Crystal Detector
Hewlett Packard Model 423A (POS), DC Power Supply HDL, Elgar, Model DCS40-30EM1-2
(13589313), and Power Amplifier JcAir (15-0777-M0).