TB 9-6625-2337-24
Use the arrow keys to highlight M4:.
(f)
ENTR, (WORD C).
(g)
Press FUNC, .
(h)
5, ENTR (TIMING, VIDEO LEVELS).
(i)
Use the arrow keys to highlight CHALLENGE VIDEO:.
(j)
5, ENTR (5.0 V).
(k)
Press FUNC, .
(l)
16, ENTR (GENERAL MEASUREMENTS).
(m)
Use the arrow keys to highlight SOURCE: ENTR (LOW PWR).
(n)
Use the arrow keys to highlight PULSE#:.
(o)
10, ENTR.
(p)
Use the arrow keys to highlight TO PULSE#:.
(q)
36, ENTR.
(r)
(3) Set signal generator for an output of 1030 MHz at +10 dBm with external
pulse modulation.
(4) Use the arrow keys to highlight SPACING: and press the ENTR key.
(5) Adjust the oscilloscope channel 1 volts/div and vertical position controls for 2 to
4 divisions of vertical deflection with the signal baseline on the center graticule line.
(6) Adjust the oscilloscope horizontal sec/div and position control to display 36 pulses.
52