TB 9-6625-2340-24
(h) SAVE, 2.
(46) Adjust A19R41 (fig. 6) for 6.750 0.001 V dc (R).
(47) Press TI keys as listed in (a) and (b) below:
(a) 2-22 GHz.
(b) RECALL, 2.
(48) Adjust A19R43 (fig. 6) for 15.750 0.001 V dc (R).
(49) Press TI keys as listed in (a) and (b) below:
(a) 2-22 GHz.
(b) SHIFT, RES BW.
(50) Disconnect multimeter from A19TP2 and A19TP3 (fig. 6).
(51) Connect a jumper between A20TP5 and A21TP2 (fig. 7).
(52) Connect microwave frequency counter Input 1 to TI 1ST LO OUTPUT through
attenuator No. 2.
(53) Set attenuator No. 2 for 10 dB.
(54) Remove 8 cable (gray) at A11J1 (fig. 7).
(55) Press TI keys as listed in (a) through (e) below:
(b) FREQUENCY SPAN, 0, Hz.
(c) SAVE, 1.
(e) SAVE, 2.
(56) Press TI RECALL, 1 and adjust A20R34 (fig. 7) for a microwave counter
indication of 2300.0 MHz 0.1 MHz (R).
(57) Press TI RECALL, 2 and adjust A20R25 (fig. 7) for a microwave counter
indication of 6150.0 MHz 0.1 MHz (R).
(58) Repeat (56) and (57) above until requirements are met.
(59) Disconnect microwave frequency counter from TI 1ST LO OUTPUT. Reconnect
termination to TI 1ST LO OUTPUT.
13