TB 9-6625-2340-24
(35) Press TI SCALE LOG key.
(36) Connect TI RF INPUT to synthesizer/level generator 50
OUTPUT.
level of 5 dBm.
(38) Adjust A4A3C55 (fig. 25) for a maximum multimeter indication.
NOTE
If A4A3C55 is at an extreme of its adjustment range (fully
meshed or unmeshed) then TI must be repaired.
(39) Short A4A3TP7 to A4A3TP8 (fig. 25) and record multimeter indication.
(40) Remove short from between A4A3TP7 and A4A3TP8.