TB 9-6625-2340-35
(11) If TI center frequency readout does not fall within limits listed in the first row of
(12) Repeat technique of (8) through (10) above for remaining center frequencies and
frequency spans listed in table 4. If TI center frequency readout does not indicate within
limits specified in table 4, perform b below.
Signal generator
Test instrument
Output
Center
frequency
frequency
Frequency
Min
Max
(GHz)
(GHz)
span
(GHz)
(GHz)
2
2
1 MHz
1.99998
2.00002
2
2
10 MHz
1.9997
2.0003
2
2
100 MHz
1.998
2.002
2
2
1 GHz
1.98
2.02
3
3
1 MHz
2.99998
3.00002
3
3
10
MHz
2.9997
3.0003
3
3
100 MHz
2.998
3.002
3
3
1 GHz
2.98
3.02
6
6
1 MHz
5.99998
6.00002
6
6
10 MHz
5.9997
6.0003
6
6
100 MHz
5.998
6.002
6
6
1 GHz
5.98
6.02
9
9
1 MHz
8.99998
9.00002
9
9
10 MHz
8.9997
9.0003
9
9
100 MHz
8.998
9.002
9
9
1 GHz
8.98
9.02
9
9
10 GHz
8.8
9.2
12
12
1 MHz
11.99998
12.00002
12
12
10 MHz
11.9997
12.0003
12
12
100 MHz
11.998
12.002
12
12
1 GHz
11.98
12.02
12
12
10 GHz
11.8
12.2
15
15
1 MHz
14.99998
15.00002
15
15
10 MHz
14.9997
15.0003
15
15
100 MHz
14.998
15.002
15
15
1 GHz
14.98
15.02
15
15
10 GHz
14.8
15.2
18
18
1 MHz
17.99998
18.00002
18
18
10 MHz
17.9997
18.0003
18
18
100 MHz
17.998
18.002
18
18
1 GHz
17.98
18.02
18
18
10 GHz
17.8
18.2
(13) Disconnect TI 10 MHz OUT (rear panel) from signal generator No. 3 FREQ.
STANDARD EXT (rear panel).
b. Adjustments
(1) Disconnect TI RF INPUT from signal generator No. 3 RF OUTPUT.