TB 9-6625-2340-35
(27) Disconnect TI RF INPUT from synthesizer/level generator 50
OUTPUT.
(28) Connect sensor module, 11792A, to measuring receiver CALIBRATION RF
POWER OUTPUT.
(29) Zero and cal the sensor module.
(30) Connect equipment as shown in fig. 27.
(31) Press TI keys as listed in (a) through (f) below:
(a)
0-2.5GHz.
(b)
TRACE A CLEAR-WRITE.
(c)
(d)
FREQUENCY SPAN, 10, MHz.
(e)
SWEEP CONT.
(f)
LOG dB/DIV, 1, dB.
(33) Set up measuring receiver to measure power at 100 MHz.
(34) Adjust signal generator No. 1 OUTPUT LEVEL control for a measuring
receiver indication of -2.0 0.1 dBm.
(35) Press TI MARKER PEAK SEARCH key. The amplitude displayed will be
within limits specified in the first row table 19.