TB 9-6625-2340-35
-14 dBm.
(6) Press TI keys as listed in (a) through (c) below:
(b) FREQUENCY SPAN, 0, Hz.
(c) ATTEN, 0, dBm.
(7) Set up measuring receiver to measure RF power.
(8) Adjust signal generator No. 3 output level for a -20.00 0.05 dBm indication on
measuring receiver.
(9) Press TI keys as listed in (a) through (f) below:
REFERENCE LEVEL, 10, -dBm.
(a)
(b)
VIDEO BW, 30, Hz.
(c)
SCALE LIN.
(d)
SHIFT.
AUTOA.
(e)
(f)
MARKER NORMAL.
(10) The marker amplitude indicated on TI display should be > -30 dBm.
(11) Press TI MARKER ∆ key.
(12) Change signal generator No. 3 output level to -4 dBm.
79