TB 9-6625-2341-24
(vi) Toggle
or
to display/highlight 1:1.
(vii) (F5) SELECT ITEM.
(viii)(F1) CLOSE which will display/highlight INPUT B.
(ix) (F1) CLOSE which will display/highlight DISPLAY INPUT B.
(c) Auto to manual range by toggling (mV RANGE V) on right side, under
INPUT B pushbutton to display/highlight B 50mVDC 1:1.
(d) "Input B" Trigger source and Trigger slope by pressing (i) through (xi) below:
(i) TRIGGER.
(ii) (F1) MORE TRIGGER.
(iii) Toggle
or
to display/highlight TRIGGER SOURCE.
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight INPUT B.
(vi) (F5) SELECT ITEM.
(vii) Toggle
or
to display/highlight TRIGGER SLOPE.
(viii) (F5) SELECT ITEM.
(ix) Toggle
or
to display/highlight + SLOPE.
(x) (F5) SELECT ITEM.
(xi) (F1) CLOSE.
(e) Time range by toggling (s TIME ns) to read 10ns/Div.
(4) Set oscilloscope calibrator LEVSINE output to produce a .5 V p-p at 150 MHz.
(5) Using oscilloscope calibrator, adjust edit dial amplitude to 2.9 vertical divisions
on TI. If TI signal is not triggered, perform b below.
(6) Set oscilloscope calibrator LEVSINE output to produce a .13 V p-p at 100 MHz.
(7) Using oscilloscope calibrator, adjust edit dial amplitude to 2.1 vertical divisions
on TI. If TI signal is not triggered, perform b below.
(8) Set range by toggling (mV RANGE V) on right side, under INPUT B
pushbutton to display/highlight B 200mVDC 1:1.
(9) Set oscilloscope calibrator LEVSINE output to produce a .3 V p-p at 10 MHz.
(10) Using oscilloscope calibrator, adjust edit dial amplitude to 1.4 vertical divisions
on TI. If TI signal is not triggered, perform b below.
(11) Set oscilloscope calibrator to standby.
(13)Press pushbutton sequence as listed in (a) through (e) below:
(a) "Input A" to on by pressing (i) through (vii) below:
(i)
INPUT A.
(ii) (F1) MORE INPUT A.
(iii) Toggle
or
to display/highlight INPUT A.
15