TB 9-6625-2341-24
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight ON.
(vi) (F5) SELECT ITEM.
(vii) (F1) CLOSE.
(b) "Input B" to off by pressing (i) through (vii) below:
(i) INPUT B.
(ii) (F1) MORE INPUT B.
(iii) Toggle
or
to display/highlight INPUT B.
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight .
(vi) (F5) SELECT ITEM.
(vii) (F1) CLOSE.
(c) "Input A" probe to 1:1 by pressing (i) through (ix) below:
(i)
INPUT A.
(ii) (F1) MORE INPUT A.
(iii) Toggle
or
to display/highlight PROBE A MENU.
(iv) (F5) SELECT ITEM which will display/highlight PROBE on INPUT A.
(v) (F5) SELECT ITEM which will display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
to display/highlight 1:1.
(vi) Toggle
or
(vii) (F5) SELECT ITEM.
(viii)(F1) CLOSE which will display/highlight INPUT A.
(ix) (F1) CLOSE which will display/highlight DISPLAY INPUT A.
(d) Auto to manual range by toggling (mV RANGE V) on left side, under
INPUT A pushbutton to display/highlight A 50mVDC 1:1.
(e) "Input A" Trigger source and Trigger slope by pressing (i) through (vii) below:
(i)
TRIGGER.
(ii) (F1) MORE TRIGGER.
(iii) Toggle
or
to display/highlight TRIGGER SOURCE.
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight INPUT A.
(vi) (F5) SELECT ITEM.
(vii) (F1) CLOSE.
(14)Set oscilloscope calibrator LEVSINE output to produce a .5 V p-p at 150 MHz.
(15)Using oscilloscope calibrator, adjust edit dial amplitude to 2.9 vertical divisions
on TI. If TI signal is not triggered, perform b below.
16