TB 9-6625-2341-24
EXT.mV needs to be selected. Change to EXT.mV mode now press (F1) YES then
(F1) CLOSE).
(h) (F1) CLOSE will display/highlight RESET MENU.
(i) (F1) CLOSE.
(4) Select range settings and display by pressing pushbuttons sequence (a) through
(m) below:
(a) SCOPE METER/MAIN MENU to display/highlight EXT.mV.
(b) SUB MENU.
(c) (F1) MORE EXT.mV.
(d) Toggle
or
to display/highlight ADD MAIN READING.
(e) (F5) SELECT ITEM.
(f) Toggle
or
to display/highlight V DC.
(g) (F5) SELECT ITEM.
(h) Toggle
or
to display/highlight READINGS on DISPLAY.
(i) (F5) SELECT ITEM.
(j) Toggle
or
to display/highlight 1.
(k) (F5) SELECT ITEM.
(l) (F1) CLOSE.
(m) SCOPEMETER/MAIN MENU to display/highlight EXT.mV.
(5) RESET calibrator.
(6) Perform steps as listed in (a) through (d) below for each row in table 8.
(a) Under INPUT A pushbutton on left side, toggle (mV RANGE V) to display
TI Range settings.
(b) Set calibrator to Output volt settings.
(c) If TI Voltage indications are not within specified limit, perform b below.
(d) Set calibrator to standby.
Calibrator
Test instrument
Voltage indication
(V dc)
Output volt settings
Range settings
Min
Max
(V dc)
(V dc)
.3
300 m
298.0 m
302.0 m
3
3
2.980
3.020
(7) Select range settings by pressing pushbuttons sequence (a) through (h) below:
(a)
SUB MENU.
(b) (F1) MORE EXT.mV.
(c) Toggle
or
to display/highlight to ADD MAIN READING.
25