TB 9-6625-2341-24
(34)
Reset calibrator.
(35)
(36)
Random sampling by pressing pushbuttons sequence (a) through (c) below:
(a) Toggle
to display/highlight (
CAL 25).
or
or
(b) Set oscilloscope calibrator to produce an EDGE square wave, 1 MHz at .6 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(37)
(38)
(39
Set calibrator to produce +12 V dc.
(40)
Press (F3) METER to display/highlight METER.
NOTE
(41)
Set calibrator to standby.
(42)
Remove +12 V dc from calibrator to TI.
(43)
(44) Select linearity calibration, zeroing ranges and ohms ranges by pressing
pushbuttons sequence (a) through (f) below:
(a) Toggle
to display/highlight (
CAL 1).
or
or
(b) Press (F5) READY, then wait till display/highlight READY is not highlighted.
NOTE
constants are being set. This process can last up to 3 minutes.
(c) Toggle
to display/highlight (
CAL 2).
or
or
(d) Press (F5) READY, then wait till display/highlight READY is not highlighted.
CAL 10).
(e) Toggle
to display/highlight (
or
or
(f)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(45)
Disconnect all shorts from TI.
(46) Select INPUT A, 300 mV range and zero for open input by toggling
to
or
display/highlight ( or
CAL 3). Press (F5) READY, then wait till display/highlight
READY is not highlighted.
(47)
NOTE
The readings on the display may not show what's being applied.
35