TB 9-6625-2341-35
NOTE
If necessary, toggle left side ▲ MOVE ▼ pushbutton to center
TI display.
(c)
If oscilloscope calibrator Voltage error is not within specified limits,
perform b below.
Test instrument
Oscilloscope calibrator
Range settings
Output volt
Voltage error
(V/Div)
settings at 1 kHz
(%)
(V p-p)
Min
Max
5
m
.02
-3
3
10
m
.04
-3
3
20
m
.08
-3
3
50
m
.2
-3
3
100
m
.4
-3
3
200
m
.8
-3
3
500
m
2
-3
3
1
4
-3
3
2
8
-3
3
5
20
-3
3
10
40
-3
3
20
80
-3
3
1
100
-3
3
50
2
100
-3
3
100
1
Adjust 5820A edit amplitude output controls for 2 vertical divisions.
2
Adjust 5820A edit amplitude output controls for 1 vertical divisions.
(12)
(F1) CLOSE.
(13)
Reset oscilloscope calibrator.
b. Adjustments.
Complete adjustments are in
paragraph 16 (R).
a. Performance Check
(1) Reset by pressing pushbutton sequence (a) through (l) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle ▲ or ▼ to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle ▲ or ▼ to display/highlight RESET SCOPEMETER.
(f) (F5) SELECT ITEM.
(g) (If all Scope Meter settings have been set to original setting then (F1)
CLOSE), which should display/highlight RESET SCOPEMETER.
(h) Toggle ▲ or ▼ to display/highlight RESET SCOPE.