TB 9-6625-2341-35
NOTE
If TRIGGER LEVEL window is open then press (F1) CLOSE
which should display/highlight CONTR.
b. Adjustments. Complete adjustments are in
paragraph 16 (R).
a. Performance Check
(1) Connect equipment as shown in figure 5 below:
(2) Reset by pressing pushbuttons sequence (a) through (k) below:
(a) SCOPE METER/MAIN MENU to display SCOPE/METER/Ω/EXT.mV and etc.
(b) (F2) METER to display/highlight METER.
(c) USER OPTIONS should display MORE OPTIONS.
(d) (F1) MORE OPTION.
(e) Toggle ▲ or ▼ to display/highlight RESET MENU.
(f) (F5) SELECT ITEM.
(g) Toggle ▲ or ▼ to display/highlight RESET METER.
(h) (F5) SELECT ITEM.
(i) All meter settings have been set to original setting then (F1) CLOSE which
should display/highlight RESET METER.
(j) (F1) CLOSE should display/highlight RESET MENU.
(k) (F1) CLOSE.
(3) Set "Input A" probe to 1:1 by pressing (a) through (g) below:
(a) INPUT A.
(b) (F5) PROBE A MENU.