TB 9-6625-2341-35
Calibrator
Test instrument
Output volt settings
Range settings
Voltage indication
(V ac)
(V ac)
(V ac)
Min
Max
.250
300 m
243.5 m
256.5 m
2
3
1.945
2.055
(9) Reset calibrator.
b.
Adjustments.
Complete adjustments are in
paragraph 16 (R).
a. Performance Check
(1) Select resistance by pressing pushbutton (F3) Ω to display/highlight Ω.
(2) Select reset by pressing pushbuttons sequence (a) through (i) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle ▲ or ▼ to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle ▲ or ▼ to display/highlight RESET OHM.
(f) (F5) SELECT ITEM.
(g) (If all OHM settings have been set to original setting then press (F1)
CLOSE), which should display/highlight RESET OHM or (for original settings, OHM
needs to be selected. Change to OHM mode now press (F1) YES then (F1) CLOSE).
(h) (F1) CLOSE should display/highlight RESET MENU.
(i) (F1) CLOSE.
(3) Press SCOPE METER/MAIN to display/highlight SCOPE/METER/Ω/EXT.mV
and etc.
(4) Perform steps as listed in (a) through (c) below for each row in table 10.
(a) Under INPUT A pushbutton on left side, toggle (mV RANGE V) to display
TI Range settings.
(b) Set calibrator to Output ohm settings.
(c) If TI Ohms indications are not within specified limits, perform b below.
Calibrator
Test instrument
Output ohm settings
Range settings
Ohms indication
(Ω)
(Ω)
(Ω)
0
30
0
300
<0.5
(5) Perform steps as listed in (a) through (d) below for each row in table 11.