TB 9-6625-2341-35
(a) Under INPUT A pushbutton on left side, toggle (mV RANGE V) to display TI
Range settings.
(b) Set calibrator to Output ohm settings.
(c) Adjust calibrator percent of error dial until calibrator reads exact indication on TI.
(d) If calibrator is not within Error indications, perform b below.
Test instrument
Calibrator
Range settings
Output ohm settings
Error indications (%)
(Ω)
(Ω)
Min
Measured
Min
30
101
-5
+5
300
100
-1
+1
3k
1k
-1
+1
30 k
10 k
-1
+1
300 k
100 k2
-1
+1
3M
1M
-1
+1
30 M
10 M
-1
+1
1
Set calibrator 2 Wire COMP to on.
2
Set calibrator 2 Wire COMP to off.
(6) Reset calibrator.
(7) Disconnect all equipment.
b.
Adjustments.
Complete adjustments are in
paragraph 16 (R).
a. Refresh Flash ROMS
(1) Before doing the complete adjustments, exit program and use oscilloscope
software Refresh Utility Disk, supplied with TI, to refresh Flash ROMS.
(2) A person should be familiar with the Refresh Utility Disk before starting the
(3) Press TI ON/OFF key to OFF and remove all connections.
(4) Remove TI battery compartment cover and battery pack.
(5) Connect PM 8907/803 AC/DC ADAPTOR supplied with TI to a 115 Vac power
source.
(6) Insert PM 8907/803 AC/DC ADAPTOR low voltage plug into TI right side
POWER ADAPTOR connector.
(7) Connect equipment as shown in figure 8 below:
25