TB 9-6625-2342-24
Table 7. MODE A Horizontal Linearity Accuracy
Test instrument
Display
A AND B
MARKER
tolerance
SEC/DIV
output
(%)
setting
s/div
200
ns/div
1.0
0.2
s/div
500
ns/div
1.0
0.5
s/div
s/div
1.0
1
1
s/div
s/div
1.0
2
2
s/div
s/div
1.0
5
5
s/div
s/div
1.0
10
10
s/div
s/div
1.0
20
20
s/div
s/div
1.0
50
50
s/div
.1
ms/div
1.0
100
s/div
.2
ms/div
1.0
200
s/div
.5
ms/div
1.0
500
1
ms/div
1
ms/div
1.0
2
ms/div
2
ms/div
1.0
5
ms/div
5
ms/div
1.0
10
ms/div
10
ms/div
1.0
20
ms/div
20
ms/div
1.0
50
ms/div
50
ms/div
1.0
.1
s/div
100
ms/div
1.0
.2
s/div
200
ms/div
1.0
.5
s/div
500
ms/div
1.0
NOTE
Ensure A AND B SEC DIV settings remain locked.
Rotate HORIZONTAL A AND B SEC/DIV control for 200 ns indications on CRT.
l.
m. Select: HORIZONTAL MODE B and set: TRIGGER MODE to AUTO.
n. Select: TRIGGER MODE. Set: RUNS AFTER to ON.
o. Use technique of g through k above for settings listed in table 8.
Table 8. MODE B Horizontal Linearity Accuracy
Test instrument
Display
A AND B
MARKER
Tolerance
SEC/DIV
output
(%)
setting
s/div
200
ns/div
1.0
0.2
s/div
500
ns/div
1.0
0.5
s/div
s/div
1.0
1
1
s/div
s/div
1.0
2
2
s/div
s/div
1.0
5
5
12