TB 9-6625-2342-24
s/div
s/div
1.0
10
10
Table 8. MODE B Horizontal Linearity Accuracy - Continued
s/div
s/div
1.0
20
20
s/div
s/div
1.0
50
50
s/div
.1
ms/div
1.0
100
s/div
.2
ms/div
1.0
200
s/div
.5
ms/div
1.0
500
1
ms/div
1
ms/div
1.0
2
ms/div
2
ms/div
1.0
5
ms/div
5
ms/div
1.0
10
ms/div
10
ms/div
1.0
20
ms/div
20
ms/div
1.0
50
ms/div
50
ms/div
1.0
.1
s/div
100
ms/div
1.0
.2
s/div
200
ms/div
1.0
.5
s/div
500
ms/div
1.0
p. Set oscilloscope calibrator to standby and disconnect equipment setup.
a. Press PRGM button then press RECALL button in main menu.
b. Use arrow labeled switches to underline FPNL.
c. Press RECALL button.
d. Rotate CH 1 VOLTS/DIV knob for a 1 V indication on CRT display.
e. Rotate CH 2 VOLTS/DIV knob for a 2 V indication on CRT display.
f. Select CURSOR FUNCTION and set TIME.
g. Rotate CURSOR/DELAY control to align active cursor to 3rd vertical graticule line.
h. Select CURSOR SELECT.
i. Rotate CURSOR/DELAY control to align active cursor to 9th vertical graticule line.
j.
TI cursor time readout will indicate between 2.9700 and 3.0300 ms.
k. Select HORIZONTAL MODE B (do not change cursor position).
l.
TI cursor time readout will indicate between 2.9700 and 3.0300 ms.
m. Set all outputs to minimum and disconnect equipment setup.