TB 9-6625-2343-50
POWER
POWER SUPPLY ADJUST
(8) Connect digital multimeter to A13TP10 (fig. 21) (positive) and A13TP15 GND
BROWN
BATTERY BACKUP THRESHOLD ADJUST
(9) Measure and record dc voltage as level No.1.
(10) Move positive lead from A13TP10 to A13TP12 (fig. 21) and record dc voltage as
level No. 2.
(11) Calculate 2/5 of level No. 1 and 1/6 of level No. 2.
(12) Multiply the lowest voltage calculated in (11) above by 0.95 and record as threshold.
49