TB 9-6625-2344-35
(6) In ACQUISITION SETUP window, place cursor over # OF AVERAGES box
and left click mouse to activate percent.
(7) Using either TI keyboard or mouse, enter 16 and click OK key.
(8) In ACQUISITION SETUP window, click CLOSE key.
(9) Click SETUP menu on top of TI display.
(10) From displayed drop-down menu, click HORIZONTAL.
(11) In HORIZONTAL SETUP window, place cursor over SCALE box and left click
mouse to activate percent.
(12) Using either TI keyboard or mouse, enter 5 n.
(13) In HORIZONTAL SETUP window, click CLOSE key.
(14) Press TI channel 1 VERTICAL INPUT button to highlight 50Ω indicator.
(15) Ensure DC coupling indicator is illuminated.
knob, set vertical scale to 200 mV.
(16) Using TI channel 1 VERTICAL
of 100 mV.
(18) Click MEASURE menu on top of TI display.
(19) From displayed drop-down menu, click MARKERS.
(20) In MARKERS SETUP window, click MODE MANUAL PLACEMENT.
(21) In MARKERS SETUP window, place cursor over MARKER A Y box and left
click mouse to activate keypad.
(22) Using either TI keyboard or mouse, enter 50 m.
(23) In MARKERS SETUP window, place cursor over MARKER B Y box and left
click mouse to activate percent.
(24) Using either TI keyboard or mouse, enter 50 m.
(25) In MARKERS SETUP window, ensure channel 1 is selected for both MARKER
A and MARKER B.
(26) In MARKERS SETUP window, click CLOSE key.
(27) Markers should be present on TI screen.
(28) Press TI TRIGGER SWEEP button to illuminate TRIG `D indicator.
knob for a stable display and TI
(29) If necessary, adjust TI TRIGGER LEVEL
(30) Repeat technique of (9) through (29) above for oscilloscope calibrator outputs and
TI settings listed in table 13 below. TI will trigger at all settings.
16