TB 9-6625-2345-35
(24) TI MKR LEVEL indication will be within limits specified in first row of table 17
∆MKR below ref column.
(25) Press TI keys as listed in (a) through (c) below:
(a) BLUE SHIFT.
(b) MKR OFF.
(c) SAVE A (not illuminated).
(26) Repeat technique of (19) through (26) above for remaining TI settings in table
17. TI
MKR LEVEL indications will be within
limits specified in
table 17 ∆MKR
indication
and ∆MKR below ref columns.
Test instrument
Ref level
∆MKR indication
∆MKR below ref
setting
(dBm)
(dBm)
(dBm)
Min
Max
Min
Max
30
11.5
9.5
33.0
31.0
40
12.0
8.0
44.0
40.0
50
12.0
8.0
54.0
50.0
60
12.0
8.0
64.0
60.0
70
12.0
8.0
74.0
70.0
80
12.0
8.0
84.0
80.0
(27) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. None
a. Performance Check
(1) Connect TI CAL OUT to TI RF INPUT 50Ω.
(2) Press TI keys as listed in (a) through (l) below:
(a) BLUE SHIFT.
(b) RESET.
(c) BLUE SHIFT.
(d) FREQ, 1, 0, 0, MHZ +dBX.
(e) BLUE SHIFT.
(f) SPAN/DIV, 1, MHZ +dBX.
(g) BLUE SHIFT.
(h) REF LEVEL, 1, 8, kHz dBX.
(i) BLUE SHIFT.
(j) dB/DIV, 1, Hz dB.
(k) VIDEO FILTER WIDE (illuminated).
(l) AUTO RESOLN (not illuminated).
(3) Rotate TI RESOLUTION BANDWIDTH control to 3 MHz.