TB 9-6625-2348-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in 4 above, and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of
Equipment Required
Common name
Minimum use specifications
Manufacturer and model
(part number)
Fluke, Model 5720A (5700A/EP)
CALIBRATOR
Dc voltage:
(p/o MIS-35947), w amplifier,
Range: 0 to 1000 V
Accuracy: .012%
Fluke, 5725A/AR (5725A/AR)
Dc current:
Range: 350 A to 10 A
Accuracy: .2%
Ac voltage
Range: 0 to 1000V
Accuracy: .175%
Range: 5 nF to 1 F
Arco Electronic,
Model
SS32
Accuracy: 5 nF 0.4%, 0.05 F 0.35%,
STANDARD
(7907233)
0.5 F 0.35%, 1 F 0.75%
FUNCTION/ARBITRARY
Range: 20 Hz, 10 MHz 1 V rms
Agilent,
Model
33250A
(MIS-
45853)
Accuracy: 0.00125%
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
CHANGE 3