TB9-6625-2350-35
NOTE
TI STO/CLR button increments displayed reading and RCL
button decrements displayed reading.
(9) Press TI STO/CLR and RCL buttons as necessary for a TI display equal to
calibrator displayed output value (rounded to TI display resolution).
(10) Press TI REL pushbutton and ensure TI REL annunciator is off.
(11) Press TI 2k range pushbutton (in position).
(12) Change calibrator output to 0 Ω nominal.
(13) Press TI REL pushbutton and ensure TI REL annunciator is displayed.
(14) Change calibrator output to 1.900 kΩ nominal.
(15) Press TI STO/CLR and RCL buttons as necessary for a TI display equal to
calibrator displayed output value (rounded to TI display resolution).
(16) Press TI REL pushbutton and ensure TI REL annunciator is off.
(17) Press TI 20k range pushbutton (in position).
(18) Change calibrator output to 0 Ω nominal.
(19) Press TI REL pushbutton and ensure TI REL annunciator is displayed.
(20) Change calibrator output to 19.000 kΩ nominal.
(21) Press TI STO/CLR and RCL buttons as necessary for a TI display equal to
calibrator displayed output value (rounded to TI display resolution).
(22) Press TI REL pushbutton and ensure TI REL annunciator is off.
(23) Disconnect calibrator SENSE HI from TI Ω SENSE HI and calibrator SENSE
LO to TI Ω SENSE LO.
(24) Press TI 200k range pushbutton (in position).
(25) Change calibrator output to 0 Ω nominal.
(26) Press TI REL pushbutton and ensure TI REL annunciator is displayed.
(27) Change calibrator output to 190.000 kΩ nominal.
(28) Press TI STO/CLR and RCL buttons as necessary for a TI display equal to
calibrator displayed output value (rounded to TI display resolution).
(29) Press TI REL pushbutton and ensure TI REL annunciator is off.
(30) Press TI MΩ range pushbutton (in position).
(31) Change calibrator output to 0 Ω nominal.
(32) Press TI REL pushbutton and ensure TI REL annunciator is displayed.
(33) Change calibrator output to 1.9 MΩ nominal.
(34) Press TI STO/CLR and RCL buttons as necessary for a TI display equal to
calibrator displayed output value (rounded to TI display resolution).
(35) Press TI REL pushbutton and ensure TI REL annunciator is off.
(36) Change calibrator output to 0 Ω nominal.
12