TB 9-6625-2351-35
c. Press TI keys as listed in (1) through (3) below:
(1) INSTRUMENT STATE RECALL.
(2) [MORE].
(3) [CRT ADJ PATTERN].
d. Adjust TRACE ALIGN (rear panel) until
leftmost line of
test pattern A
(fig. 1) is
parallel with crt bezel.
e. Adjust X POSN (rear panel) until
leftmost @
characters B
(fig. 1) and
softkey labels
C (fig. 1) appear just inside left and right edges of crt bezel.
f. Adjust Y POSN (rear panel) until the
softkey labels C
(fig. 1) align with the
appropriate softkeys.
g. Press INSTRUMENT STATE PRESET key.
8. Calibrator Output Accuracy
Test
a. Performance Check
NOTE
If necessary, perform measuring receiver and sensor module
(11722A) ZERO and CALIBRATE.
(1) Connect measuring receiver sensor module input to TI CAL OUTPUT.