TB 9-6625-2352-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Analog Multimeters 0 to 6000 Volts (General) listed in table 1. The
manufacturers' manuals or TMs were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in text, tables and figures.
b. Time and Technique. The time
required for each calibration is
approximately 2
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Tables for specifications, calibration
Models or
performance limits, and figures for adjustments
Manufacturers
military designators
Tables
Figures
---
ME-297//U
3
1
---
TS-352( )/U
4
2
Simpson
260
5
---
Simpson
260-3
6
3
Simpson
260-4
7
4
Simpson
260-5
8
5
Simpson
269
9
---
Simpson
269 Series III
10
6
Simpson
270
11
7
Triplett
310
12
---
Triplett
630
13
---
Triplett
630NA
14
8
Triplett
630PL
15
---
Triplett
666R
16
---