TB 9-6625-2364-24
(a) System Inputs, [A Input Settings].
(b) More, [ChA Filter], [Filter Off On] to On, [Mode Man Auto] to Man.
(c) [Length 128], using ⇐,⇒,⇑ and ⇓ keys, set filter length to 0512 in pop up
window then press [Enter].
(6) Press TI Zero Cal, [Zero A] keys and wait for Please Wait symbol to disappear.
(7) Wait 30 seconds. TI reading will be 0 0.05 W.
(8) Press TI POWER and range calibrator LINE switches to off position.
(9) Move cable connection from TI CHANNEL A power sensor input to TI
CHANNEL B power sensor input.
(10) Set TI POWER switch to on position.
(11) Press Preset Local, [Confirm] keys.
(12) Press TI DISPLAY ↕ key to activate B display.
(13) Repeat technique of (3) (c) through (7) above for channel B.
b. Adjustments. None.
9. Instrument Accuracy
Test
a. Performance Check
(1) Press TI POWER and range calibrator LINE switches to off position.
(3) Unplug power cord from range calibrator.