TB 9-6625-2365-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Electronic Counter, Anritsu, Models MF2414B and MF2414B Option 003.
The manufacturer's manual was used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports.
a.
Forms, records and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are in table 1.
Test instrument parameters
Performance specifications
Frequency: 10 MHz
Time base:
Stability:
<2 X 10-8/day (after 24 hour warm-up)
<5 X 10-10/day (after 48 hour warm-up) (option 003)
Input 2
Frequency range:
10 Hz to 1 GHz
Accuracy:
1 count time base accuracy X measurement frequency
trigger error (10 Hz to 10 MHz)
1 count time base accuracy X measurement frequency
(10 MHz to 1 GHz)
Sensitivity: 50 Ω: 10 MHz to 1 GHz, 25 mVrms
1MΩ: 10 Hz to 10 MHz, 25 mVrms
Input 1
Frequency range:
600 MHz to 40 GHz
Accuracy:
1 count time base accuracy X measurement frequency
Residual error 1(measurement frequency (GHz)/10
count (rms))
Sensitivity:
-33 dBm (< 12.4 GHz)
-28 dBm (<20 GHz)
-25 dBm (<26.5 GHz)
dBm = 0.741F (GHz) -44.6 (≤40 GHz)
Burst mode
Frequency range:
600 MHz to 40 GHz
Accuracy:
1 count time base accuracy X measurement frequency
trigger error Residual error 2 (measurement frequency
(GHz)/2 count (rms)) 1/Tgw
Pulse width:
100 ns to 100 ms
Width
Accuracy:
20 ns time base accuracy X measurement pulse width
trigger error