TB 9-6625-2365-35
Table 2. Minimum Specifications of Equipment Required (Continued)
SYNTHESIZED SIGNAL
10 MHz to 40 GHz Anritsu, Model 68369NV
(68369NV)
Accuracy: 0.025 Hz of output
Power range:
-19 to +4 dBm
Datum, Model ET6000-75
TIME/FREQUENCY
Frequency: 10 MHz
(13589305)
WORKSTATION
Accuracy: 1.25 parts in 10-10 per day
VARIABLE
26.5 to 40 GHz
PRD, Model 157-F1
ATTENUATOR NO. 1
Accuracy:
(10519383)
1
primary
VARIABLE
26.5 to 40 GHz
PRD, Model 157-F1
ATTENUATOR NO. 2
Accuracy:
(10519383)
1
primary
WAVEGUIDE TO
26.5 to 40 GHz
Wiltron, Model 35WR28K
COAXIAL ADAPTER
Insertion loss: 0.5 dB (maximum)
(13335442-3)
NO. 1
(WR28 TO K MALE)1
WAVEGUIDE TO
26.5 to 40 GHz
Wiltron, Model 35WR28K
COAXIAL ADAPTER
Insertion loss: 0.5 dB (maximum)
(13335442-3)
NO. 2
(WR28 TO K FE MALE)1
1Limited deployed.
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Press TI power line switch (rear panel) to on (detent position).