TB 9-6625-2802-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Paper speed
Range:
1, 5, 20, 100 mm/SEC
Accuracy: + 0.5%
Span accuracy Range:
1 mv/DIV to 5v/DIV
Accuracy: + 0.5%
Linearity
Range:
2 mv/DIV, 50 mv/DIV, 2v/DIV
Accuracy: + 0.5 mm of decrement
Step response Input:
5 volt peak-to-peak square wave at
10 Hz.
Accuracy: Less than 2% overshoot
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific
exceed the minimum use specifications listed in table 2.
equipment used in this calibration procedure. This
The accuracy's listed in table 2 provide a four-to-one
equipment is issued with the secondary transfer
accuracy ratio between the standard and TI.
standards calibration set, NSN 6695-00621-7877 and is
to be used in performing this procedure. Alternate items
5. Accessories Required. The accessories listed in
may be used by the calibrating activity when the
equipment listed in table 2 is not available. The items
selected must be verified to perform satisfactorily prior to
necessary, these items may be substituted by equivalent
items unless specifically prohibited.
equipment must meet or
3