TB 9-6625-357-35
Figure 3. Model 3202 (R) adjustment -
top view
k. Adjust P206 (fig. 3) for minimum dc
voltage indication on
multimeter while turning
CUTOFF FREQ Hz dial from 200 to 20.
l. Repeat f and h above.
16. Noise Level and Stability
NOTE
repeat for right section.
a. Performance Check
(1) Set CUTOFF FREQ Hz dial to 20 and selector switch to HIGH PASS X1.
(2) Short INPUT connector.
Noise level, as indicated on
NOTE
All TI covers must be in place. It may be necessary to use
isolation plug on multimeter.