TB 9-6635-203-35
7. Equipment Setup
a. Remove TI and calibration bar from carrying case.
b. Compare serial number on name plate of TI with serial number stamped on tab of
c. Rotate cable size-gage ccw against stop pin
(fig. 1).
Figure 1. Dial indicating tensiometer (typical).
d. Turn bezel until indicator dial indicates 1/16.
NOTE
This setting must be changed to correspond to size of cable
being used.
e. Compress TI
actuating handle and
insert calibration bar into
jaws with
calibration
f. Release actuating handle completely.