TB 9-6670-258-35
(a) Set power switch to 115 V 60~.
(b) Operate BATTERY VOLTAGE switch on indicator panel of TI and adjust AC
VOLTAGE ADJUST control until pointer of balance meter indicates in green area of
milliammeter scale.
(3) If 24 V dc is used as power source, operate BATTERY VOLTAGE switch and
check polarity. If meter indication is in yellow area, change leads and operate BATTERY
VOLTAGE switch again. If meter indication is in red area, battery voltage is either too
high or too low and should be replaced.
(4) Allow sufficient time for equipment to warm-up and stabilize.
(5) Set up equipment as shown in figure 5. Use load cell No. 4 with compatible
spherical button and TI cell No. 1 (red).
Figure 5. System equipment setup -
alternate technique (reference).
(6) Bolt LOWER ADAPTER (fig. 6) to bottom of load cell No. 4. Mount on LOWER
10