TB 9-6695-242-24
(5) Zero load cell indicator.
a. Performance Check.
Slowly apply tension (force) for TI indications listed in
NOTE
Final value of tension must be approached using increasing
values. If desired point is overshot, decrease tension several
hundred pounds and again increase tension to desired point.
Table 3. Accuracy for Dillon, Model AN-ELEX
Load cell indictor
Test instrument
indications
indications
(Lbs)
(Lbs)
Min
Max
3000
2550
3450
4000
3550
4450
5000
4550
5450
6000
5550
6450
7000
6550
7450
8000
7550
8450
9000
8550
9450
10,000
9550
10,450
11,000
10,550
11,450
12,000
11,550
12,450
13,000
12,500
13,450
14,000
13,500
14,450
15,500
14,550
15,450
Table 4. Accuracy for Dillon, Model AN-IONO
Load cell indicator
Test instrument
indications
indications
(Lbs)
(Lbs)
Min
Max
2000
1600
2400
4000
3600
4400
6000
5600
6400
8000
7600
8400
10,000
9600
10,400
12,000
11,600
12,400
14,000
13,600
14,400
16,000
15,600
16,400
18,000
17,600
18,400
20,000
19,600
20,400