TB 9-6695-272-35
NOTE
It may be necessary to add thermal converter correctional data
for frequency readings that are not within tolerance, or
interrelate for frequency not checked on thermal converter
calibration report. Allow thermal converter to settle before
readings are recorded.
(2) Set FREQUENCY to 100 kHz and press +13.01 dBm.
(4) Press +12.91 dBm.
(6) Calculate the allowable tolerance by subtracting reading in (5) above from
reading in (3) above.
(7) Record tolerance.
(8) Reset amplitude to +13.01 dBm.
(9) Set TI to the following frequencies, allow multimeter reading to settle and record
the reading: 200 Hz, 50 kHz, 500 kHz, 1 MHz, 5 MHz, and 10 MHz (see NOTE above).
(10) The result for each frequency must indicate the reading of (3) above + the
tolerance established in (7) above.
(11) Disconnect figure 4, CONNECTION A, equipment and reconnect as shown in
10