TB 9-6695-288-35
(2) Position controls as listed in (a) through (f) below:
(a) FUNCTION switch to INPUT LEVEL.
(b) FILTERS pushbuttons to off (out).
(c) DISTORTION RANGE switch to 0 dB.
(d) INPUT RANGE switch to .1 V.
(e) INPUT/GND SELECT switch to DIS.AN/1 (center position).
(3) Set oscillator to 10 kHz and output level for an indication of 1 V on TI meter.
and output amplitude to 42 dBm.
(5) Set FUNCTION switch to DISTORTION.
(6) Set DISTORTION RANGE switch to -80 dB.
(7) Adjust function/arbitrary generator amplitude control as necessary to obtain a
distortion reading of 80 dB on TI meter (full scale indication).
16 CHANGE 1
PIN: 063104-001