TB 9-6695-293-50
Test instrument
parameters
Performance specifications
Calibrator, John Fluke, Model 5700A/( ) AND 5720A/( )
Range: 10 Hz to 30 MHz
Accuracy: 0.01%
Wideband ac voltage
Voltage output:
(Option 03) (cont.)
Absolute uncertainty: (% of output + V)3
Range: 1.1 mV (- 46 dBm) to 3.5 V (+ 24 dBm)
Frequency: 10 Hz to 30 MHz
Accuracy:
Frequency
Range
1.1 mV
3.0 mV
11 mV
33 mV
- 46 dBm
- 37 dBm
- 26 dBm
- 17 dBm
(% output + V)
10 - 30 Hz
1.1 + 2
1.0 + 3
1.0 + 8
0.9 + 16
30 Hz - 500 kHz
0.8 + 2
0.7 + 3
0.7 + 8
0.6 + 16
0.5 - 2.0 MHz
1.0 + 5
0.8 + 6
0.8 + 11
0.7 + 19
2.0 - 10 MHz
1.2 + 5
1.1 + 6
0.9 + 11
0.8 + 19
10 - 20 MHz
1.4 + 5
1.3 + 6
1.1 + 11
1.0 + 19
20 - 30 MHz
2.3 +
2.2 + 18
1.7 + 11
1.6 + 19
17
110 mV
330 mV
1.1 V
3.5 V
- 6.2 dBm
+ 3.4 dBm + 14 dBm
+ 24 dBm
(% output + V)
10 - 30 Hz
0.9 + 40
0.8 + 100
0.8 + 400
0.7 + 500
30 Hz - 500 kHz
0.6 + 40
0.5 + 100
0.5 + 400
0.4 + 500
0.5 - 2.0 MHz
0.7 + 43
0.6 + 103
0.6 + 403
0.5 + 503
2.0 - 10 MHz
0.8 + 43
0.7 + 103
0.7 + 403
0.6 + 503
10 - 20 MHz
1.0 + 43
0.9 + 103
0.9 + 403
0.8 + 503
20 - 30 MHz
1.6 + 43
1.5 + 103
1.5 + 403
1.4 + 503
Amplifier, John Fluke 5725A/( )
Ac voltage
Output: 110 to 1100 V ac
Accuracy: (ppm of output + mV)
Range
Frequency
40 Hz to 1 kHz
1 to 20 kHz
20 to 30 kHz
1100
90 + 4
165 + 6
600 + 11
Range
Frequency
30 to 50 kHz
50 to 100 kHz
750
600 + 11
0.23 % + 45
Dc current
Range: 11 A
Accuracy: (360 ppm of output + 480 A)
Ac current
Range : 11 A
Accuracy: (ppm of output + A)
Frequency
40 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
460 + 170
950 + 380
0.36 % + 750
See footnotes at end of table.
6