TB 9-6695-293-50
NOTE
When turned on, the TI undergoes a self test. If self test fails,
take corrective action before proceeding.
c. Perform the DC ZERO CALIBRATI0N by pressing the following softkeys:
(1) SETUP MENUS
(2) CAL
(3) ZERO
NOTE
Through the following calibration steps the TI display screen
will provide some instructions. When the display reads
"...Please Wait...," wait until display changes before continuing
with next step.
a. Press RESET pushbutton then press SETUP MENUS softkey.
b. Press CAL softkey; then press CAL softkey again.
c.
Press one of the
softkeys under the
CALIBRATION display.
d. Enter ambient temperature and press ENTER.
f. Enter the test report value of the dc reference standard 10 V output; then press
ENTER and wait for PROCEED to appear on TI display.