Paragraphs b through d below comprise the verification test
for the wideband option. Paragraph e below is the flatness
calibration procedure. Performance of e below is only required
if an out-of-tolerance condition exists in b and d below or upon
paragraphs a above and c below must be completed prior to d
and e below in order to establish, or characterize, voltage
accuracies at 1 kHz. If an out-of-tolerance condition exists in c
below, repeat a above and c below. If an out-of-tolerance
condition still exists, a repair action is required.
b. Frequency Accuracy
(1) Connect wideband output to frequency counter CHANNEL A input, using
termination and cable supplied with TI. Set TI for a 0 dBm, 10 Hz wideband output.
(2) Measure output frequency; frequency counter will indicate between 99.99 and
100.01 ms.
(3) Adjust TI frequency to those listed in table 9. At each frequency setting repeat
(2) above. Frequency counter will indicate within the listed limits.
(4) Press RESET pushbutton.
Test instrument
Frequency counter indication
frequency
Min
Max
100 Hz
9.999
ms
10.001
ms
300 Hz
299.97 Hz
300.03 Hz
500 Hz
499.95 Hz
500.05 Hz
800 Hz
799.92 Hz
800.08 Hz
900 Hz
899.91 Hz
900.09 Hz
1
kHz
999.9 Hz
1.0001
kHz
1.19 kHz
1.189881 kHz
1.190119 kHz
2.2 MHz
2.19978 MHz
2.20022 MHz
3.5 MHz
3.49965 MHz
3.50035 MHz
3.8 MHz
3.79962 MHz
3.80038 MHz
10
MHz
9.999
MHz
10.001
MHz
20
MHz
19.998
MHz
20.002
MHz
30
MHz
29.997
MHz
30.003
MHz
27