TB 9-6695-293-50
Figure 20. John Fluke, Model 5220A/CT -
board location.
NOTE
The operation of the TI is controlled by the calibrator. The TI
controls referred to in the remainder of this check are located
on the calibrator.
30. Dc Accuracy and Line
Regulation
a. Performance Check
(1) Replace standard resistor No. 3 (.1Ω) with standard resistor No. 4 (.01Ω).
(2) Set TI for a +10.0 A dc output.