TB 9-6695-298-35
WARNING
HIGH VOLTAGE is used or exposed during the performance
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Set power switch on
rear panel of TI to the
off position if it is not already
off.
b. Disconnect any transducers that may be connect to TI.
c. Remove top cover of TI (four screws on top and two on each side).
d. Install
connector plug P1 into J1
(fig. 1) and P2 into J2
(fig. 1) of TI if not
already
installed.
e. Connect the strain
gage simulator to CHANNEL 1
(J1, fig. 2) of TI and
set it to 0
percent.
f. Insert EEPROM
module into MODULE COMPARTMENT
(fig. 3) (if it is not
installed).
g. Replace top cover on TI.
h. Connect 115V ac
power to
105-130 VAC
50/60 Hz FUSE 2A
(fig. 2) power input on
rear panel of TI.
i. Set power switch on
rear panel of TI to the on
position and allow 30 minutes for
warmup.