TB 9-6695-301-50
On some models of CALIBRATOR, the handwheel is replaced by an electric motor.
1
If your CALIBRATOR does not have sufficient space to connect the LOAD CELL and the TEST INSTRUMENT
2
together in the lower assembly, replace the SPACE BLOCK in the upper assembly with the LOAD CELL.
Figure 2. Load cell tension check -
equipment setup.
(2) Place a mild steel pad on top of load button of standard load cell. Use tension
members and adapters as needed.
CAUTION
Do not exceed range of standard load cell or TI load cell.
NOTE
Zero out tare weight of all suspended items (tension).
(3) Connect standard load cell indicator to standard load cell and TI indicator to TI
load cell. Connect power cables to 115 V ac power source. Energize all equipment and
allow 15 minute warm up.