TB 11-6625-2715-50
SECTION IV
CALIBRATION
setup and adjust current shunt for 30 amp reading.
(8) Adjust current supply for a TI full scale
(1) Adjust Current Supply A3 and Decade
indication.
Resistor A4 for a TI full scale indication.
(2) Observe that the DC Differential Voltmeter
indicates between 294 and 30.6 millivolts.
indicates between 14.7 and 15.3 millivolts.
b. Adjustment. No adjustments can be made.
(3) Remove positive lead from 1.5 ampere
connector and connect to TI 3.0 ampere connector.
1-10. Final Procedure. a. Deenergize and disconnect
(4) Adjust current supply and decade resistor
all equipment.
for a TI full scale indication.
b. In accordance with TM 38-750, annotate and
indicates between 24 and 30.6 millivolts.
the TI cannot be adjusted within tolerance, annotate and
(6) Remove positive lead from 3.0 ampere
affix Form 2417 (Unserviceable or Limited Use) (Red
connector and connect to TI 30 ampere connector.
Tag).
(7) Remove decade resistor from equipment
5