TB 55-4920-346-34
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin provides instructions for the calibration of
environment control test set,
Grumman, Model A/E 24T-102. The manufacturer's manual was used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique.
The time required for this calibration is approximately 6 hours, using the DC
technique.
2. DA Form 2416 (Calibration Data Card).
a. Forms, records, and reports required for calibration personnel at all levels are prescribed by TB 750-25. DA
Form 2416 must be annotated in accordance with TB 750-25 for each calibration performed.
b. Adjustments to be reported on DA Form 2416 are designated (R) at the end of the sentence in which they
appear. When adjustments are in tables, the (R) follows the designated adjustment. Report only those adjustments
made and designated with (R).
3. Calibration Description.
TI parameters and performance specifications which pertain to this calibration are listed in
table 1.
Table 1. Calibration Description
Test Instrument Parameters
Performance Specifications
Input voltages
115 vac
10 v. 400 Hz 20 HZ
Inductors
12.25 vac 0.25 v
Meter phase voltage
110 vac
2.0 v
Milliamps
0-360 ma ac 5.0 FS
0-500 ma dc 5.0 FS
Temperature simulator
0-99 K ohms 0.5%
Sensor simulator
0-99 K ohms 0.5%
2